Quantum Scientific Imaging - QSI 583ws 8.3mp Cooled CCD Camera w/Mechanical Shutter & CFW - DISCONTINUED

Model#: 583ws

Discontinued Product

Manufactured by: Quantum Scientific Imaging, Inc.

Details

This is a discontinued item. We recommend the QSI 683ws Cooled CCD Camera with Mechancial Shutter and CFW as a replacement.

QSI 583ws Cooled CCD Camera

The QSI 583 camera employs an 8.3 megapixel Kodak Enhanced Response full-frame CCD image sensor with microlens technology. The high quantum efficiency, wide dynamic range and low noise performance make the 532 ideally suited to a broad range of demanding astronomical, scientific and industrial imaging applications.

The KAF-8300 sensor has a photoactive array of 3326W x 2504H pixels. It has excellent quantum efficiency between 350nm and 1000nm with significant enhancement at the blue end of the spectrum. Low dark current and high pixel charge capacity result in a dynamic range exceeding 70db. The sensor employs a true two-phase charge transfer technology with a transparent gate that significantly increases optical response compared to traditional front illuminated full frame sensors. Micro lenses cover the surface of the CCD and focus the light through the transparent gate to further increase optical response.


Mechanical Shutter

The Model 583s incorporates the internal mechanical shutter in the 'mid-size' camera body configuration . The shutter is not only used for timing exposures, but can be left closed to produce 'dark frames' for subsequent image processing. Exposures can range from 0.03 seconds to 240 minutes.


583ws adds Internal Color Filter Wheel

The Model 583ws adds an internal five position filter wheel to the camera in a 'full-size' camera body. The filter wheel accepts any standard 1.25" filter and can be purchased with various filter configurations. The filter wheel can easily be removed and replaced to change or clean the glass filters. Additional filter wheels can be purchased allowing quick interchange of different filter set configurations.

The close positioning of the internal filter wheel allows unvignetted images with scopes as fast as f/4 using standard 1.25" filters, or as fast as f/2.8 utilizing the new unmounted 31mm filters. Save hundreds of dollars over cameras that require external filter wheels with larger diameter filters.


583wsg adds Integrated Guider Port

The Model 583wsg camera has the internal mechanical shutter and color filter wheel like the 583ws, plus it adds an Integrated Guider Port. This allows guiding using the light from your main telescope while picking off the light from the guide star in front of the filters.

 


The Benefits of Scientific Imaging Performance

QSI's scientific-grade CCD cameras deliver image quality equal to or better than cameras costing thousands more. The lower noise and other benefits of a QSI camera allow you to capture better data in less time. What does scientific imaging performance really mean?  There are four key traits that distinguish a scientific camera:

  1. High sensitivity
  2. Lower Noise
  3. Linear response to light
  4. Precisely regulated cooling

 

High Sensitivity 

How sensitive a CCD it to light is expressed as its Quantum Efficiency (QE).  Higher QE means you can collect more photons in any given amount of time giving you better looking images with a higher Signal to Noise Ratio (SNR).  High sensitivity is almost purely a function of the CCD.  Just about any camera using the same CCD will achieve a similar sensitivity to light. 

However, even with the same CCD, the Signal to Noise Ratio between two different cameras can be dramatically different especially in the fainter, lower signal areas of an image.  These lower signal areas, such as the fringes of nebulae or dust lanes in a galaxy, often comprise some of the most interesting parts of astronomical images.  Maximizing the SNR in these dim areas is the key to revealing those otherwise hidden details.

Lower Noise = Higher Dynamic Range 

Noise in a CCD camera is uncertainty in the signal you’re trying to measure.  The lower the noise, the greater the dynamic range possible with any given CCD.  There are several types of noise that contribute to the overall noise in a CCD image.  

CCD images are exquisitely sensitive to any noise introduced by the camera’s electronics.  Mitigating the electronic interference from the digital electronics and switching power supplies is one of the many challenges in designing and building a scientific grade CCD camera.  Any noise introduced by the camera is relatively easy to see by analyzing bias frames.  A Bias Frame is a zero-length (dark) exposure which captures the slightly varying “floor value” from pixel to pixel (the bias) plus any noise added during the process of reading the image from the CCD.  By subtracting two bias frames, you can remove the underlying differences in the pixel bias leaving just the noise added to the image when it was read from the CCD.  This subtraction of two bias frames is called a Read Noise Frame

Complicating the story is that some types of noise are tougher to deal with than others.  For instance, truly random noise that displays a Gaussian distribution can be reduced by combining images.  The graph to the right shows the histogram of a bias frame from a QSI camera overlaid over histograms from two other cameras using the same Kodak CCD. The histograms have been normalized for easy visual comparison. The histogram from the QSI camera displays a nearly perfect Gaussian distribution. The familiarbell curve resembles a parabola when displayed with a log scale. Standard deviation (σ) is shown for each histogram.

The parabolic shape of the histogram is important but you can't infer too much about the magnitude of the noise from the width of a histogram without knowing more about the cameras. Differences in width can be caused by higher noise or possibly just differences in a camera's gain settings.

Averaging together multiple bias frames can tell you more about the distribution of pixel values and make it easier to visually identify any non-Gaussian noise in the images.

To the left is a histogram comparing the average of 10 bias frames from a QSI 520i to a camera from another manufacturer with the same sensor taken at the same -10C temperature. The graph has the same scale as the one above for the single bias frames.

Note that both histograms are now narrower than the histograms for the single frames. This is a result of the noise being reduced in the combined frames because the random (Gaussian) noise tends to cancel out getting you closer to the "true" value of the underlying data. Also note however that the red histogram trails off to the right no longer resembling a parabola like the blue histogram from the QSI camera. This non-Gaussian noise typically has components of both pattern noise and stray pixels that fall outside the target Gaussian distribution. Pattern noise in particular will not be reduced by combinging frames. Instead this type of noise combines just like signal reinforcing itself and reducing the signal to noise ratio in your final images.

Bias frame histograms, however, can’t tell you the full story.  The width and shape of the histogram shows the magnitude and overall distribution of the noise, but it tells you little about the structure of that noise.  To do that requires another tool, the Fourier Transform.

A Fourier Transform is a mathematical function that converts a signal (any signal) into a collection of sine waves which together exactly represent the original signal.  For images, we’re interested in 2-dimensional Fourier Transforms which convert an image from the Spatial Domain, rows and columns in a digital picture, into an equivalent Frequency Domain image. A Fast Fourier Transform (FFT) is a special algorithm that runs much faster on computers when starting with square images where each side has a length equal to a power of 2 (e.g 256, 512, 1024).

To analyze the noise from a CCD camera, you create a Read Noise Frame as described above. By subtracting two bias frames, you end up with just the noise introduced by reading the image from the CCD. An FFT of that Read Noise Frame will reveal any pattern noise introduced by the camera. 

An FFT of random noise displaying a Gaussian distribution will have a smooth dark background with a single bright pixel in the center. To the left is an FFT of a Read Noise Frame from a QSI 500 Series camera. The FFT of the QSI Read noise frame reveals no pattern noise and is, in fact, virtually indistinguishable from an FFT of mathematically generated Gaussian noise.

To the right is the FFT of a Read Noise Frame from a camera that introduces non-Gaussian, periodic noise into the image.  Note the series of vertical lines running through the FFT and the three bright dots near the center.  Those vertical lines in the FFT are the result of horizontal noise of various frequencies in the original image.

Periodic noise will not be reduced when combining multiple images; instead it is reinforced, just like signal. The net effect of such noise is to raise the noise floor, reducing the signal to noise ratio and hiding the faint details in your images.  When many such images are combined, periodic noise may show up as visible patterns, such as faint horizontal or vertical lines, running through your images.

Roll your mouse over the image to the right to see an example from a camera that displays even stronger pattern noise.

The images below of a standard optical target demonstrate one of the benefits of low noise and a Gaussian distribution.  The images were captured with a QSI 516 camera in a very dark room.  A single image closely resembles a bias frame. The target is virtually indistinguishable from the background.  Its Signal to Noise Ratio is nearly zero.  With an average combine of 50 frames, light and dark areas are clearly visible.  With 200 frames, the target pattern is easily identifiable.  Note that even after combining 200 frames, the difference between the dark and bright pixels is only 0.45 ADUs on average.

Lower noise displaying a Gaussian distribution allows you to reveal
subtle details in your targets that would otherwise be lost


 

Linear Response

Accurately recording the relative amount of light between different objects requires a camera that delivers a linear response to varying levels of light.  A CCD camera should record twice the photon flux from an object that is twice as bright.  Cameras that use CCDs with anti-blooming protection will start to show a drop in linearity as the pixels begin to approach a high percentage of full well capacity, but should be nearly linear at lower signal levels.

 

The log scale graph above shows the response to light of a QSI 520i which uses an interline transfer CCD with anti-blooming protection. the log scale of the graph amplifies any discrepencies from linear response especially at the lowest signal levels.   The data points are linear to less then 1% all the way down to the lowest signal level, well within the specification of the CCD.  Lower noise is especially critical in maintaining linearity at these lower signal levels. 

High linearity is obviously critical for scientific applications such as accurate photometry, but it's just as important for producing beautiful pictures of deep sky objects. A linear response to light and lower noise allows you to accurately capture the subtle details of your targets in less time.  To achieve the same signal to noise ratio in the faintest areas of your images with a higher noise camera can require hours more exposure time.  Turning that around, a lower noise camera will yield great looking images with a higher signal to noise ratio in much less time.

Precisely Regulated Cooling

Every image taken with a CCD camera has two distinct types of “signal”.  A CCD records light striking it from the objects in your field of view, and it records heat from the sensor itself.  Cooling the CCD reduces the contribution of this “thermal” current.  With most Kodak CCDs, the amount of thermal current is cut in half for every roughly 6°C drop in the temperature of the CCD.  So, dropping the temperature of the CCD by 36°C reduces the rate that thermal current builds by over 98%.

The rate that thermal current builds at any given temperature is very predictable and repeatable.  That’s why dark frame subtraction is so effective.  Subtracting an average or median combine of multiple dark frames taken at the same temperature as your light frames will remove the thermal current component (and bias) from your light frames.  However, if the temperature varies either during the capturing of your light frames or dark frames by even a few tenths of a degree, dark frame subtraction becomes much less effective.

 

The graph above plots the temperature of the CCD in a QSI 500 Series camera cooling from an ambient temperature of 13.5°C (56°F) down to -20°C. The drop of a little over 33°C below the ambient temperature took just over six minutes. QSI 500 Series cameras will generally cool the CCD to approximately 38°C below the ambient temperature and then regulate the temperature to within 0.1°C.This fine control of the temperature allows for very precise and very repeatable dark frame subtraction. 600 Series cameras have increased cooling capacity to approximately 45°C below the ambient temperature.

Summary

High sensitivity, low noise, a linear response to light and precisely regulated cooling are four key traits of a scientific camera.  All are required in order to maximize the dynamic range of your images.  Given a constant signal, lower noise increases the Signal to Noise Ratio and dynamic range, allowing you to produce better images in less time.  Capturing good looking images of bright targets is relatively easy, but if you want great results in the dimmer areas of your targets, achieving the same signal to noise ratio with a higher noise camera can require hours of additional dark time.

All significant performance characteristics of QSI cameras, including LinearityRead Noise, and Photon Transfer (Gain), are tested and documented as the final manufacturing step. Each camera's timing and voltages are carefully set to ensure maximum Charge Transfer Efficiency, and to minimize charge injection and other secondary noise sources.

QSI cameras have many other benefits including small size, minimal backfocus, available internal color filter wheel, low power consumption and exceptional build quality.

 

 

 

 

 

 


Product Series

The QSI 500 Series is a family of medium format, thermo-electrically cooled CCD cameras designed to produce high quality images with extremely wide dynamic range, excellent linearity and exceptionally low noise. Seven different models are available employing a comprehensive range of monochrom and single-shot color scientific grade CCDs. A variety of options and accessories are available to meet your specific astronomical, medical or industrial imaging objectives.

The superb imaging performance of 500 Series cameras is wrapped in an attractive, compact design with outstanding power efficiency and investment protecting upgradeability. The refined engineering and impressive fit, finish and attention to detail will surpass your highest expectations.

 

 

 

Additional Information

Name Quantum Scientific Imaging - QSI 583ws 8.3mp Cooled CCD Camera w/Mechanical Shutter & CFW - DISCONTINUED
Manufacturer Quantum Scientific Imaging, Inc.
Model 583ws
Discountinued Yes
Antiblooming 1000x suppression
Computer Interface USB 2.0 (USB 1.1 compatible)
Cooling Temperature regulation +/- 0.1°C, @ 0°C to -40°C CCD temperature
Dimensions 4.45" x 4.45" x 2.5"
Full Frame Download Typically <21 second (host computer dependent)
Mount Standard adapter - T-Thread, 42mm x .75mm
Objective Nosepiece Standard, T-Adapter to 2" nosepiece
Sensor Type Kodak KAF-8300
Shutter Mechanical, exposure range: 0.03 seconds to 240 minutes